This software application enables to upload firmware to Neurobit devices. It is also possible to read out information about the hardware and firmware from the device. Wireless (Bluetooth or IrDA) or USB communication with the equipment is used. The application works under Microsoft® Windows® operating systems: 11/10/8/7/Vista/XP.
This information appears after connection to a device:
Neurobit Optima[+] USB devices:
A cable with full size USB 2.0 type A plug on one end and micro B plug
on the other end (the same as for typical cell phone chargers) is
required. Install driver software for USB chip by FTDI, as described in the
device quick start guide. Connect the device to the computer and turn it on.
The computer should detect the device and install necessary USB driver.
Neurobit Optima[+] BLE devices:
Windows 8.1 or newer system is required.
Bluetooth Low Energy communication port (built-in into a computer or as a USB adapter)
compatible with the protocol version 4.2 or newer is necessary.
To install Bluetooth adapter follow Neurobit Optima Quick Start Guide.
Neurobit Optima device should be set in wireless communication range (usually
a few meters). Before the use a BLE device should be paired with your computer.
Pairing code for Neurobit Optima[+] is 000000. You can learn more about Bluetooth pairing
in Windows help.
Neurobit Optima[+] BT and older Neurobit Optima (without link type suffix) devices:
Bluetooth communication port (built-in into a computer or as a USB adapter)
compatible with the protocol version 2.0 or newer is required. To install
Bluetooth adapter follow Neurobit Optima Quick Start Guide.
Neurobit Optima device should be set in wireless communication range (usually
a few meters). Before the use a BT device should be paired with your computer.
Pairing code for Neurobit Optima[+] is 0000. You can learn more about Bluetooth pairing
in Windows help.
Neurobit Lite devices:
Infrared (IR) communication port (built-in into a computer or in form of a USB adapter)
compatible with IrDA standard v. 1.0 or newer is required. To install IrDA adapter
follow an instruction delivered in its package.
The Neurobit Lite device should be set with IR port oriented towards IR port
of your computer, no more than 0.5m from it, without physical obstructions
between both entities, preferably on the same axis. There should be no other
active IrDA devices nearby.
Download the latest firmware binary:
Run Winload program:
Click Browse to select downloaded firmware file and choose Link type used by your Neurobit device. (If necessary, write down the Device id.)
Turn on Neurobit device (and wait for its beep signal).
Click Upload button of the application. Information about the hardware and its firmware is displayed. The bar at the bottom of the window indicates firmware transmission progress. The upload usually takes less than a dozen seconds for Bluetooth and USB.
After successful transmission the new firmware
is automatically launched. It is signalled with a beep.
(In Neurobit Lite
the new firmware name and version will be displayed on device's screen.)
You can now turn off the device and close the loader application.
If there are several Neurobit Optima[+] devices paired with /
connected to your computer, we recommend to turn off all of them but the one
intended for firmware update.
(Otherwise the updated device should be
unambiguously identified with its transceiver identifier set in the
Device id.)
If a firmware transmission failed for any reason (like loss of connection between the devices or heavy load of processor), try to upload the firmware again.
If you plan to use several Neurobit devices (having the same link type) with the same computer concurrently, you need to identify each device unambiguously with its transceiver identifier. To be able to do so follow the procedure at the beginning:
From now on you can use these identifiers to specify, which device piece will be used in a given instance.
If Neurobit Lite configuration structure has changed in new firmware version,
default configuration is set. It is indicated with an appropriate message on
device's display.
In this case please make again startup settings, as described in the chapter
"The basics of usage/Initial configuration" of your equipment instruction
manual (especially System/Mains option, according to your country).
You may also want to set other configuration options then, depending on your
preferences.
For most updates the device's configuration is preserved.
You can also consult the chapter "Firmware update" of device instruction manual.